COMMISSION IMPLEMENTING REGULATION (EU) 2022/1107
of 4 July 2022
laying down common specifications for certain class D
in vitro
diagnostic medical devices in accordance with Regulation (EU) 2017/746 of the European Parliament and of the Council
(Text with EEA relevance)
Article 1
Common specifications
Article 2
Definitions
Article 3
Transitional provisions
Article 4
Entry into force and date of application
ANNEX I
GENERAL COMMON SPECIFICATIONS
Part I – Requirements for performance characteristics of devices covered by Annexes II to XIII
Performance characteristics |
Requirement |
||||||||||||||||
All performance characteristics set out in Section 9.1, points (a) and (b), Section 9.3 and Section 9.4, point (a), of Annex I to Regulation (EU) 2017/746 |
|
||||||||||||||||
Whole system failure rate |
|
||||||||||||||||
Analytical sensitivity and analytical specificity, interference |
|
||||||||||||||||
Analytical and diagnostic specificity, interference and cross-reactivity |
|
||||||||||||||||
Batch-to-batch consistency |
|
Part II – Requirements for performance characteristics of devices referred to in Annexes III to XIII
Performance characteristic |
Requirement |
||||||||||||||
Analytical and diagnostic sensitivity |
|
||||||||||||||
Analytical and diagnostic specificity |
|
||||||||||||||
Analytical and diagnostic specificity, interference and cross-reactivity |
|
||||||||||||||
Performances obtained by lay persons |
|
ANNEX II
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OF BLOOD GROUP ANTIGENS IN THE ABO, RH, KELL, DUFFY AND KIDD BLOOD GROUP SYSTEMS
Scope
Reagent specificity |
Number of tests per method claimed by the manufacturer |
Total number of specimens to be tested for a launch device |
Total number of specimens to be tested for a new formulation, or use of well-characterised reagents |
General qualification criteria |
Specific qualification criteria |
Acceptance criteria |
Anti-ABO1 (Anti-A), Anti-ABO2 (Anti-B), Anti-ABO3 (Anti-A,B) |
≥500 |
≥3 000 |
≥1 000 |
Clinical specimens: 10 % of the test population Neonatal specimens: > 2 % of the test population |
ABO specimens shall include > 40 % A and B antigen positive specimens which may include specimens from group A, group B and group AB |
All of the reagents shall show comparable performance to state-of-the-art CE marked devices with regard to claimed reactivity of the device. For CE marked devices where the application or use has been changed or extended, further testing shall be carried out in accordance with the requirements outlined in column 2 above (“Number of tests per method claimed by the manufacturer”). |
Anti-RH1 (Anti-D) |
≥500 |
≥3 000 |
≥1 000 |
Performance evaluation of Anti-D reagents shall include tests against a range of weak RH1 (D) and partial RH1 (D) specimens, depending on the intended use of the product. Weak and/or partial D cells shall account for > 2 % of RH1 (D) positive specimens. |
||
Anti-RH2 (Anti-C), Anti-RH4 (Anti-c), Anti- RH3 (Anti-E) |
≥100 |
≥1 000 |
≥200 |
|
||
Anti-RH5 (Anti-e) |
≥100 |
≥500 |
≥200 |
|
||
Anti-KEL1 (Anti-K) |
≥100 |
≥500 |
≥200 |
|
||
Anti-JK1 (Jka), Anti-JK2 (Jkb) |
≥100 |
≥500 |
≥200 |
|
||
Anti-FY1 (Fya), Anti-FY2 (Fyb) |
≥100 |
≥500 |
≥200 |
|
||
Note: Positive specimens used in the performance evaluation shall be selected to reflect variant and weak antigen expression. |
Table 2. Manufacturer’s batch-to-batch consistency testing of reagents and reagent products to determine blood group antigens in the ABO, Rh, Kell, Duffy and Kidd blood group systems
1.
Test reagents
Blood group reagents |
Minimum number of control cells to be tested as part of specificity testing |
Acceptance criteria |
|||||||
|
Positive reactions |
|
Negative reactions |
Each batch of reagent shall exhibit unequivocal positive or negative results by all techniques claimed by the manufacturer in accordance with the results obtained from the performance evaluation data. |
|||||
|
A1 |
A2B |
Ax |
|
B |
O |
|
||
Anti-ABO1(Anti-A) |
2 |
2 |
2(1) |
|
2 |
2 |
|
||
|
B |
A1B |
|
|
A1 |
O |
|
||
Anti-ABO2(Anti-B) |
2 |
2 |
|
|
2 |
2 |
|
||
|
A1 |
A2 |
Ax |
B |
O |
|
|
||
Anti-ABO3(Anti-A,B) |
2 |
2 |
2(1) |
2 |
4 |
|
|
||
|
R1r |
R2r |
WeakD |
|
r’r |
r”r |
rr |
||
Anti-RH1 (Anti-D) |
2 |
2 |
2(1) |
|
1 |
1 |
1 |
||
|
R1R2 |
R1r |
r’r |
|
R2R2 |
r”r |
rr |
||
Anti-RH2 (Anti-C) |
2 |
1 |
1 |
|
1 |
1 |
1 |
||
|
R1R2 |
R1r |
r’r |
|
R1R1 |
|
|
||
Anti-RH4 (Anti-c) |
1 |
2 |
1 |
|
3 |
|
|
||
|
R1R2 |
R2r |
r”r |
|
R1R1 |
r’r |
rr |
||
Anti-RH3 (Anti-E) |
2 |
1 |
1 |
|
1 |
1 |
1 |
||
|
R1R2 |
R2r |
r”r |
|
R2R2 |
|
|
||
Anti-RH5 (Anti-e) |
2 |
1 |
1 |
|
3 |
|
|
||
|
Kk |
|
|
|
kk |
|
|
||
Anti-KEL1 (Anti-K) |
4 |
|
|
|
3 |
|
|
||
|
Jk(a+b+) |
|
|
|
|
Jk(a–b+) |
|
|
|
Anti-JK1 (Anti-Jka) |
4 |
|
|
|
|
3 |
|
|
|
|
Jk(a+b+) |
|
|
|
|
Jk(a+b–) |
|
|
|
Anti-JK2 (Anti-Jkb) |
4 |
|
|
|
|
3 |
|
|
|
|
Fy(a+b+) |
|
|
|
|
Fy(a–b+) |
|
|
|
Anti-FY1 (Anti-Fya) |
4 |
|
|
|
|
3 |
|
|
|
|
Fy(a+b+) |
|
|
|
|
Fy(a+b–) |
|
|
|
Anti-FY2 (Anti-Fyb) |
4 |
|
|
|
|
3 |
|
|
|
Note: Polyclonal reagents shall be tested against a wider panel of cells to confirm specificity and exclude presence of unwanted contaminating antibodies. |
2.
Control materials (red cells)
ANNEX III
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HUMAN IMMUNODEFICIENCY VIRUS (HIV) INFECTION
Scope
Definitions
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 HIV-1 ≥100 HIV-2 including 40 non-B-subtypes including 25 positive ‘same day’ fresh serum specimens (≤ 1 day after specimen taking) all available HIV/1 subtypes shall be represented by at least 3 specimens per subtype |
all true positive specimens shall be identified as positive |
|
Seroconversion panels |
≥30 panels at least 40 early seroconversion HIV specimens shall be tested |
diagnostic sensitivity during seroconversion shall represent the state of the art all seroconversion HIV specimens shall be identified as positive |
Diagnostic specificity |
Unselected blood donors (including first-time donors)(1) |
≥5 000 |
≥99,5% |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total (such as RF+, from related virus infections, from pregnant women, subjects recently vaccinated against any infectious agent) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 HIV-1 ≥100 HIV-2 including 40 non-B-subtypes all available HIV/1 subtypes shall be represented by at least 3 specimens per subtype |
all true positive specimens shall be identified as positive |
Seroconversion panels |
≥30 panels at least 40 early seroconversion HIV specimens shall be tested |
diagnostic sensitivity during seroconversion shall represent the state of the art all seroconversion HIV specimens shall be identified as positive |
|
Diagnostic specificity |
Unselected blood donors (including first-time donors) |
≥1 000 |
≥ 99 % |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥200 specimens from pregnant women ≥100 other potentially cross-reacting specimens in total (e.g. RF+, from related infections) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥200 HIV-1 ≥100 HIV-2 Including different stages of infection and reflecting different antibody patterns |
Identification as “confirmed positive” or “indeterminate”, not as “negative” |
Seroconversion panels |
≥15 seroconversion panels/low titre panels ≥40 early seroconversion HIV specimens |
Diagnostic sensitivity during seroconversion shall represent the state of the art All seroconversion HIV specimens shall be identified as positive |
|
Diagnostic specificity |
Blood donors |
≥200 |
No false positive results / no neutralisation |
Hospitalised patients |
≥200 |
||
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 in total (including specimens from pregnant women, specimens with indeterminate results in other confirmatory assays) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥50 HIV-1 antigen positive ≥50 cell culture supernatants including different HIV-1 subtypes and HIV-2 |
all true positive specimens shall be identified as positive (after neutralisation if applicable) |
Seroconversion panels |
≥20 seroconversion panels/low titre panels ≥40 early seroconversion HIV specimens |
diagnostic sensitivity during seroconversion shall represent the state of the art all seroconversion HIV specimens shall be identified as positive |
|
Analytical sensitivity |
First International Reference Reagent HIV-1 p24 Antigen, NIBSC code: 90/636 |
|
≤ 2 IU/ml |
Diagnostic specificity |
Blood donors |
≥200 |
≥ 99,5 % after neutralisation or, if no neutralisation test available, after resolution of the specimen status |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 |
Table 5. Qualitative and quantitative NAT devices for HIV RNA
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Analytical sensitivity |
WHO International Standard HIV-1 RNA; WHO International Standard HIV-2 RNA; or calibrated reference materials |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimum 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device. LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).(2) quantitative NAT: definition of lower, upper quantification limit, precision, accuracy, ‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels |
According to the state of the art |
HIV geno-/subtype sensitivity |
all relevant genotypes/subtypes, preferably from international reference materials potential substitutes for rare HIV subtypes (to be quantified by appropriate methods): cell culture supernatants; in vitro transcripts; plasmids |
Qualitative NAT: at least 10 specimens/genotype or subtype Quantitative NAT: dilution series for demonstration of quantification efficiencies |
According to the state of the art |
Diagnostic sensitivity |
Positive specimens reflecting the routine conditions of users (e.g. no pre-selection of specimens) |
Quantitative NAT: ≥100 Comparative results with another NAT system shall be generated in parallel |
According to the state of the art |
Seroconversion panels |
Qualitative NAT: ≥10 panels Comparative results with another NAT system shall be generated in parallel |
According to the state of the art |
|
Diagnostic specificity |
Blood donor specimens |
Qualitative NAT: ≥500 Quantitative NAT: ≥100 |
According to the state of the art |
Cross-reactivity |
Potentially cross-reacting specimens |
≥10 human retrovirus positive specimens (e.g. HTLV) |
According to the state of the art |
Carry-over |
High HIV RNA positive; HIV RNA negative |
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. |
According to the state of the art |
Detection in relation to antibody status |
HIV-RNA positives: anti-HIV negative, anti-HIV positive |
Pre-seroconversion (anti-HIV negative) and post-seroconversion (anti-HIV positive) specimens |
According to the state of the art |
Whole system failure rate |
HIV RNA low-positive |
≥100 HIV RNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration. |
≥99% positive |
Performance characteristic |
Specimens(3) |
Number of lay persons |
||||||||
Result interpretation(4) |
Interpretation of results(5) by lay persons reflecting the following range of reactivity levels:
|
≥ 100 |
||||||||
Diagnostic sensitivity |
lay persons that are known positive |
≥ 200 |
||||||||
Diagnostic specificity |
lay persons that do not know their status |
≥ 400 |
||||||||
Lay persons that are at high risk of acquiring the infection |
≥ 200 |
ANNEX IV
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HUMAN T-CELL LYMPHOTROPIC VIRUS (HTLV) INFECTION
Scope
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
|
|||
Diagnostic sensitivity |
Positive specimens |
≥300 HTLV-I ≥100 HTLV-II including 25 positive ‘same day’ fresh serum specimens (≤ 1 day after specimen taking) |
all true positive specimens shall be identified as positive |
Seroconversion panels |
To be defined when available |
diagnostic sensitivity during seroconversion shall represent the state of the art, if applicable |
|
Diagnostic specificity |
Unselected blood donors (including first-time donors)(1) |
≥5 000 |
≥ 99,5% |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total (e.g. RF+, from related virus infections, from pregnant women) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥300 HTLV-I ≥100 HTLV-II |
all true positive specimens shall be identified as positive |
Seroconversion panels |
To be defined when available |
diagnostic sensitivity during seroconversion shall represent the state of the art, if applicable |
|
Diagnostic specificity |
Unselected blood donors (including first-time donors) |
≥1 000 |
≥ 99% |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥200 specimens from pregnant women ≥100 other potentially cross-reacting specimens in total (e.g. RF+, from related infections) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥200 HTLV I ≥100 HTLV II |
Identification as “confirmed positive” or “indeterminate”, not as “negative” |
Seroconversion panels |
To be defined when available |
diagnostic sensitivity during seroconversion shall represent the state of the art, if applicable |
|
Diagnostic specificity |
Blood donors |
≥200 |
No false positive results |
Hospitalised patients |
≥200 |
||
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 in total (including specimens from pregnant women, specimens with indeterminate results in other confirmatory assays) |
Table 4. NAT devices for HTLV I/II
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Analytical sensitivity |
International reference preparations |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimum 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device. LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).(2) quantitative NAT: definition of lower, upper quantification limit, precision, accuracy, ‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels |
According to the state of the art |
HTLV I and HTLV II genotype sensitivity |
all relevant genotypes, preferably from international reference materials potential substitutes for rare HTLV genotypes (to be quantified by appropriate methods): cell culture supernatants; in vitro transcripts; plasmids |
Qualitative NAT: at least 10 specimens/genotype or subtype Quantitative NAT: dilution series for demonstration of quantification efficiencies |
According to the state of the art |
Diagnostic specificity |
Blood donor specimens |
Qualitative NAT: ≥500 Quantitative NAT: ≥100 |
According to the state of the art |
Cross-reactivity |
Potentially cross-reacting specimens |
≥10 human retrovirus positive specimens (e.g. HIV-1, HIV-2) |
According to the state of the art |
Carry-over |
High HTLV RNA positive; HTLV RNA negative |
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. |
According to the state of the art |
Detection in relation to antibody status |
HTLV-RNA positives: anti-HTLV negative, anti-HTLV positive |
Pre-seroconversion (anti-HTLV negative) and post-seroconversion (anti-HTLV positive) specimens |
According to the state of the art |
Whole system failure rate |
HTLV RNA low-positive |
≥100 HTLV RNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration. |
≥99% positive |
ANNEX V
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HEPATITIS C VIRUS (HCV) INFECTION
Scope
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 Including specimens from different stages of infection and reflecting different antibody patterns HCV genotype 1-4: > 20 specimens per genotype (including non-a subtypes of genotype 4); HCV genotypes 5 and 6: > 5 specimens each; including 25 positive ‘same day’ fresh serum specimens (≤ 1 day after specimen taking) |
all true positive specimens shall be identified as positive |
|
Seroconversion panels |
≥30 panels HCV seroconversion panels for the evaluation of HCV antigen and antibody combined tests (HCV Ag/Ab) shall start with one or more negative bleeds and comprise panel members from early HCV infection (HCV core antigen and/or HCV RNA positive but anti-HCV negative). |
diagnostic sensitivity during seroconversion shall represent the state of the art HCV Ag/Ab tests shall demonstrate enhanced sensitivity in early HCV infection when compared to HCV antibody only tests. |
Diagnostic specificity |
Unselected blood donors (including first-time donors)(1) |
≥5 000 |
≥99,5% |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total (e.g. RF+, from related virus infections, from pregnant women) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 including specimens from different stages of infection and reflecting different antibody patterns. HCV genotype 1-4: > 20 specimens per genotype (including non-a subtypes of genotype 4); HCV genotypes 5 and 6: > 5 specimens each; |
all true positive specimens shall be identified as positive |
Seroconversion panels |
≥30 panels HCV seroconversion panels for the evaluation of HCV antigen and antibody combined tests (HCV Ag/Ab) shall start with one or more negative bleeds and comprise panel members from early HCV infection (HCV core antigen and/or HCV RNA positive but anti-HCV negative). |
diagnostic sensitivity during seroconversion shall represent the state of the art HCV Ag/Ab tests shall demonstrate enhanced sensitivity in early HCV infection when compared to HCV antibody only tests. |
|
Diagnostic specificity |
Unselected blood donors (including first-time donors)1 |
≥1 000 |
≥ 99 % |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥200 specimens from pregnant women ≥100 other potentially cross-reacting specimens in total (e.g. RF+, from related infections) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥300 Including specimens from different stages of infection and reflecting different antibody patterns. HCV genotypes 1 – 4: > 20 specimens (including non-a subtypes of genotype 4; HCV genotype 5: > 5 specimens; HCV genotype 6: as far as available |
identification as “confirmed positive” or “indeterminate”, not as “negative” |
Seroconversion panels |
≥15 seroconversion panels/low titre panels |
diagnostic sensitivity during seroconversion shall represent the state of the art |
|
Diagnostic specificity |
Blood donors |
≥200 |
No false positive results/ no neutralisation |
Hospitalised patients |
≥200 |
||
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 in total (including specimens from pregnant women, specimens with indeterminate results in other confirmatory assays) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥25 HCV core antigen and/or HCV RNA positive but anti-HCV negative specimens, comprising HCV genotypes 1-6 (if a genotype is not available, a justification shall be made) |
all true positive specimens shall be identified as positive |
Seroconversion panels |
≥20 seroconversion panels/low titre panels HCV seroconversion panels for the evaluation of HCV antigen and antibody combined tests shall start with one or more negative bleeds and comprise panel members from early HCV infection (HCV core antigen and/or HCV RNA positive but anti-HCV negative). |
diagnostic sensitivity during seroconversion shall represent the state of the art HCV antigen and antibody combined tests shall demonstrate enhanced sensitivity in early HCV infection when compared to HCV antibody only tests. |
|
Analytical sensitivity |
WHO International Standard HCV core (PEI 129096/12) |
Dilution series |
|
Diagnostic specificity |
Blood donors |
≥200 |
≥ 99,5 % after neutralisation or, if no neutralisation test available, after resolution of the specimen status |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 |
Table 5. Qualitative and quantitative NAT devices for HCV RNA
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Analytical sensitivity |
WHO International Standard HCV RNA(or calibrated reference materials) |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimum 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device. LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).(2) quantitative NAT: definition of lower, upper quantification limit, precision, accuracy, ‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels |
According to the state of the art |
HCV genotype sensitivity |
all relevant genotypes/subtypes, preferably from international reference materials potential substitutes for rare HCV genotypes (to be quantified by appropriate methods): in vitro transcripts; plasmids |
Qualitative NAT: ≥10 specimens/genotype or subtype Quantitative NAT: dilution series for demonstration of quantification efficiencies |
According to the state of the art |
Diagnostic sensitivity |
Positive specimens reflecting the routine conditions of users (e.g. no pre-selection of specimens) |
Quantitative NAT: ≥100 Comparative results with another NAT system shall be generated in parallel |
According to the state of the art |
Seroconversion panels |
Qualitative NAT: ≥10 panels Comparative results with another NAT system shall be generated in parallel |
According to the state of the art |
|
Diagnostic specificity |
Blood donor specimens |
Qualitative NAT: ≥500 Quantitative NAT: ≥100 |
According to the state of the art |
Cross-reactivity |
Potentially cross-reacting specimens |
>10 human flavivirus (e.g. HGV, YFV) positive specimens |
According to the state of the art |
Carry-over |
High HCV RNA positive; HCV RNA negative |
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. |
According to the state of the art |
Detection in relation to antibody status |
HCV RNA positives: anti-HCV negative, anti-HCV positive |
Pre-seroconversion (anti-HCV negative) and post-seroconversion (anti-HCV positive) specimens |
According to the state of the art |
Whole system failure rate |
HCV RNA low-positive |
≥100 HCV RNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration. |
≥99% positive |
Performance characteristic |
Specimens(3) |
Number of lay persons |
||||||||
Result interpretation(4) |
Interpretation of results(5) by lay persons reflecting the following range of reactivity levels:
|
≥ 100 |
||||||||
Diagnostic sensitivity |
lay persons that are known positive |
≥ 200 |
||||||||
Diagnostic specificity |
lay persons that do not know their status |
≥ 400 |
||||||||
lay persons that are at high risk of acquiring the infection |
≥ 200 |
ANNEX VI
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HEPATITIS B VIRUS (HBV) INFECTION
Scope
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 anti-HBc: including evaluation of different HBV markers HBsAg: including different HBV genotypes / subtypes / mutants anti-HBc or HBsAg: including 25 positive ‘same day’ fresh serum (≤ 1 day after specimen taking) |
Overall performance shall be at least equivalent to the comparator device |
Seroconversion panels |
HBsAg assays: ≥30 panels anti-HBc assays: to be defined when available |
diagnostic sensitivity during seroconversion shall represent the state of the art (for anti-HBc this shall be the case if applicable) |
|
Analytical sensitivity |
WHO Third International Standard HBsAg (subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226) |
|
For HBsAg assays: <0,130 IU/ml |
Diagnostic specificity |
Unselected blood donors (including first-time donors)(1) |
≥5 000 |
≥99,5% |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total (e.g. RF+, from related virus infections, from pregnant women,) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 including evaluation of different HBV markers including different HBV genotypes / subtypes / mutants |
Overall performance shall be at least equivalent to that of the comparator device |
Seroconversion panels |
HBsAg assays: ≥30 panels anti-HBc assays: to be defined when available |
Diagnostic sensitivity during seroconversion shall represent the state of the art (for anti-HBc this shall be the case if applicable) |
|
Diagnostic specificity |
Unselected blood donors (including first-time donors) |
≥1 000 |
HBsAg assays: ≥ 99 % anti-HBc assays: ≥ 99 % |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥200 specimens from pregnant women ≥100 other potentially cross-reacting specimens in total (e.g. RF+, from related infections) |
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥300 Including specimens from different stages of infection Including 20 ‘high positive’ specimens (>26 IU/ml); 20 specimens in the cut-off range |
Correct identification as positive (or indeterminate), not negative |
Seroconversion panels |
≥15 seroconversion panels/low titre panels |
Diagnostic sensitivity during seroconversion shall represent the state of the art |
|
Analytical sensitivity |
WHO Third International Standard for HBsAg, subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226 |
|
|
Diagnostic specificity |
Negative specimens |
≥10 false positives as available from the performance evaluation of the first-line assay |
No false positive results/ no neutralisation |
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 |
Performance characteristic |
|
anti-HBs |
anti-HBc IgM |
anti-HBe |
HBeAg |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥100 vaccinees ≥100 naturally infected persons |
≥200 Including specimens from different stages of infection (acute/chronic, etc.) |
≥200 Including specimens from different stages of infection (acute/chronic, etc.) |
≥200 Including specimens from different stages of infection (acute/chronic, etc.) |
≥ 98 % (for anti-HBc IgM: applicable only on specimens from acute infection stage) |
Seroconversion panels |
10 anti-HBs seroconversion panels or follow-up series |
When available |
When available |
When available |
Diagnostic sensitivity during seroconversion shall represent the state of the art (for anti-HBc IgM, anti-HBe, HBeAg this shall be the case if applicable) |
|
Analytical sensitivity |
Standards |
WHO Second International Standard for anti-hepatitis B surface antigen (anti-HBs) immunoglobulin, human NIBSC code: 07/164 |
|
WHO First International Standard anti-hepatitis B virus e antigen (anti-HBe), PEI code 129095/12 |
WHO First International Standard for Hepatitis B Virus e Antigen (HBeAg) PEI code 129097/12 HBe |
anti-HBs: < 10 mIU/ml |
Diagnostic specificity |
Negative specimens |
≥500 Including clinical specimens ≥50 potentially interfering specimens |
≥200 blood donations ≥200 clinical specimens ≥50 potentially interfering specimens |
≥200 blood donations ≥200 clinical specimens ≥50 potentially interfering specimens |
≥200 blood donations ≥200 clinical specimens ≥50 potentially interfering specimens |
≥ 98 % |
Table 5. Qualitative and quantitative NAT devices for HBV DNA
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Analytical sensitivity |
WHO International Standard HBV DNA (or calibrated reference materials) |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimum 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device. LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).(2) quantitative NAT: definition of lower, upper quantification limit, precision, accuracy, ‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels |
According to the state of the art |
HBV genotype sensitivity |
WHO International Reference Panel HBV DNA (HBV genotypes) all relevant genotypes/subtypes, preferably from international reference materials potential substitutes for rare HBV genotypes (to be quantified by appropriate methods): plasmids; synthetic DNA |
Qualitative NAT: at least 10 specimens/genotype or subtype Quantitative NAT: dilution series for demonstration of quantification efficiencies |
According to the state of the art |
Diagnostic sensitivity |
Positive specimens reflecting the routine conditions of users (no pre-selection of specimens) |
Quantitative NAT: ≥100 Comparative results with another NAT system shall be generated in parallel |
According to the state of the art |
Seroconversion panels |
Qualitative NAT: ≥10 panels Comparative results with another NAT system shall be generated in parallel |
According to the state of the art |
|
Diagnostic specificity |
Blood donor specimens |
Qualitative NAT: ≥500 Quantitative NAT: ≥100 |
According to the state of the art |
Cross-reactivity |
Potentially cross-reacting specimens |
|
According to the state of the art |
Carry-over |
High HBV DNA positive; HBV DNA negative |
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. |
According to the state of the art |
Detection in relation to antibody status |
HBV DNA positives: anti-HBV negative, anti-HBV positive |
Pre-seroconversion (anti-HBV negative) and post-seroconversion (anti-HBV positive) specimens |
According to the state of the art |
Whole system failure rate |
HBV DNA low-positive |
≥100 HBV DNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration. |
≥99% positive |
Performance characteristic |
Specimens(3) |
Number of lay persons |
||||||||
Result interpretation(4) |
Interpretation of results(5) by lay persons reflecting the following range of reactivity levels:
|
≥100 |
||||||||
Diagnostic sensitivity |
lay persons that are known positive |
≥200 |
||||||||
Diagnostic specificity |
lay persons that do not know their status |
≥400 |
||||||||
lay persons that are at high risk of acquiring the infection |
≥200 |
ANNEX VII
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HEPATITIS D VIRUS (HDV) INFECTION
Scope
Performance characteristic |
|
anti-HDV |
anti-HDV IgM |
Delta antigen |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥100 Specifying markers of HBV coinfection |
≥50 Specifying markers of HBV coinfection |
≥10 Specifying markers of HBV coinfection |
≥ 98 % |
Diagnostic specificity |
Negative specimens |
≥200 Including clinical specimens ≥50 potentially interfering specimens |
≥200 Including clinical specimens ≥50 potentially interfering specimens |
≥200 Including clinical specimens ≥50 potentially interfering specimens |
≥ 98 % |
Table 2. Qualitative and quantitative NAT devices for HDV RNA
Performance characteristic |
Specimen |
Specimen numbers, features, use |
Acceptance criteria |
Analytical sensitivity |
WHO First International Standard HDV RNA, PEI code 7657/12 |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimum 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device. LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).(1) quantitative NAT: definition of lower, upper quantification limit, precision, accuracy, ‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels |
According to the state of the art |
HDV genotype sensitivity |
all relevant genotypes/subtypes, preferably from international reference materials potential substitutes for rare HDV genotypes (to be quantified by appropriate methods): plasmids; synthetic RNA |
Quantitative NAT: dilution series for demonstration of quantification efficiencies |
According to the state of the art |
Diagnostic specificity |
Blood donor specimens |
Qualitative NAT: ≥100 Quantitative NAT: ≥100 |
According to the state of the art |
Cross-reactivity |
Potentially cross-reacting specimens |
|
According to the state of the art |
Carry-over |
High HDV RNA positive; HDV RNA negative |
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. |
According to the state of the art |
Whole system failure rate |
HDV RNA low-positive |
≥100 HDV RNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration. |
≥99% positive |
ANNEX VIII
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OF MARKERS OF VARIANT CREUTZFELDT-JACOB (vCJD) DISEASE
Scope
Performance characteristic |
Material |
Number of specimens |
Acceptance criteria |
||
Analytical sensitivity |
vCJD brain spikes in human plasma (WHO reference number NHBY0/0003) |
≥24 replicates of each of three dilutions of the material WHO number NHBY0/0003 (1×104, 1×105, 1×106) |
23 of the 24 replicates detected at 1×104 |
||
vCJD spleen spikes in human plasma (10% spleen homogenate — NIBSC reference number NHSY0/0009) |
≥24 replicates of each of three dilutions of the material NIBSC number NHSY0/0009 (1×10, 1×102 , 1×103 ) |
23 of the 24 replicates detected at 1×10 |
|||
Diagnostic sensitivity |
Specimens from appropriate animal models |
As many specimens as reasonably possible and available, and ≥10 specimens |
90% |
||
Specimens from humans with known clinical vCJD |
As many specimens as reasonably possible and available, and ≥10 specimens |
90% |
|||
Only in cases where 10 specimens are not available:
|
max. one false negative result |
||||
Analytical specificity |
Potentially cross-reacting specimens |
≥100 |
|
||
Diagnostic specificity |
Normal human plasma specimens from area of low bovine spongiform encephalopathy (BSE) exposure |
≥5 000 |
≥ 99,5 % |
ANNEX IX
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF CYTOMEGALOVIRUS (CMV) INFECTION
Scope
Performance characteristic |
Specimens |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 including specimens from recent and past CMV infection, low and high positive titre specimens |
≥ 99% sensitivity for confirmable past infection(1); overall sensitivity including recent infection(2) shall be at least equivalent to the comparator device |
Seroconversion panels |
To be tested when available |
Diagnostic sensitivity during seroconversion shall represent the state of the art |
|
Analytical sensitivity |
Standards |
WHO International Standard anti-CMV IgG (PEI-code 136616/17) In case of titre determinations and quantitative statements |
|
Diagnostic specificity |
Negative specimens |
≥400(3) CMV negative specimens from unselected donors, as compared to another CMV test. |
≥ 99% |
Hospitalised patients(4) |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting(5) specimens |
≥100 in total (e.g. RF+, related viruses or other infectious agents, pregnant women, etc.) |
Table 2. Qualitative and quantitative NAT devices for CMV DNA
Performance characteristic |
Specimens |
Specimen numbers, features, use |
Acceptance criteria |
Analytical sensitivity |
WHO First International Standard Human CMV DNA (09/162; 5 000 000 IU/vial) (or calibrated reference materials) |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimum 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device. LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).(6) quantitative NAT: definition of lower, upper quantification limit, precision, accuracy, ‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels |
According to the state of the art |
Diagnostic sensitivity CMV Strain sensitivity |
Patient specimens determined as CMV DNA positive by comparator device Dilution series of CMV positive cell cultures may serve as potential substitutes |
Qualitative NAT: ≥100 Quantitative NAT: ≥100 dilution series for demonstration of quantification efficiencies |
According to the state of the art |
Diagnostic specificity |
Blood donor specimens |
Qualitative NAT: ≥500 Quantitative NAT: ≥100 |
According to the state of the art |
Cross-reactivity |
Potentially cross-reacting specimens |
≥20 specimens in total Including human specimens positive for related human herpesviruses, e.g. EBV, HHV6, VZV Herpesvirus positive cell cultures may serve as potential substitutes |
According to the state of the art |
Carry-over |
High CMV DNA positive; CMV DNA negative |
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. |
According to the state of the art |
Whole system failure rate |
CMV DNA low-positive |
≥100 CMV DNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration. |
≥99% positive |
ANNEX X
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF EPSTEIN-BARR VIRUS (EBV) INFECTION
Scope
Performance characteristic |
Specimens |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 including specimens from recent and past EBV infection, low and high positive titre specimens |
≥ 99% for confirmable past infection(1); overall sensitivity including recent infection(2) shall be at least equivalent to the comparator device |
Seroconversion panels |
To be tested when available |
diagnostic sensitivity during seroconversion shall represent the state of the art |
|
Analytical sensitivity |
Standards |
International reference reagents, when available |
|
Diagnostic specificity |
Negative specimens |
≥ 200(3) EBV negatives from unselected donors as compared to another EBV device |
≥ 99% |
Hospitalised patients(4) |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total (e.g. RF+, related viruses or other infectious agents, pregnant women, etc.) |
Table 2. Qualitative and quantitative NAT devices for EBV DNA
Performance characteristic |
Specimens |
Specimen numbers, features, use |
Acceptance criteria |
Analytical sensitivity |
WHO First International Standard Human EBV DNA (09/260; 5 000 000 IU/vial) (or calibrated reference materials) |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimum 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device. LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).(5) quantitative NAT: definition of lower, upper quantification limit, precision, accuracy, ‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels |
According to the state of the art |
Diagnostic sensitivity EBV strain sensitivity |
Patient specimens determined as EBV DNA positive by comparator device Dilution series of EBV positive cell cultures may serve as potential substitutes |
Qualitative NAT: ≥100 Quantitative NAT: ≥100 dilution series for demonstration of quantification efficiencies |
|
Diagnostic specificity |
Negative specimens |
Qualitative NAT: ≥500 Quantitative NAT: ≥100 |
According to the state of the art |
Cross-reactivity |
Potentially cross-reacting specimens |
≥20 specimens in total Including human specimens positive for related human herpesviruses, e.g. CMV, HHV6, VZV Herpesvirus positive cell cultures may serve as potential substitutes |
According to the state of the art |
Carry-over |
High EBV DNA positive; EBV DNA negative |
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. |
According to the state of the art |
Whole system failure rate |
EBV DNA low-positive |
≥100 EBV DNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration. |
≥99% positive |
ANNEX XI
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OF MARKERS OF
TREPONEMA PALLIDUM
INFECTION
Scope
Performance characteristic |
Specimens |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥200 positive specimens in total, at different stages of the infection if available, including high positive and low positive specimens, identified as positive by at least two different serological tests (one of which is an enzyme immunoassay) for different antibodies to T.pallidum |
≥99.5% overall sensitivity |
Seroconversion panels |
At least 1 seroconversion panel, ≥1 if possible, including individual specimens from the early infection phase |
Diagnostic sensitivity during seroconversion shall represent the state of the art. |
|
Analytical sensitivity |
Standards |
WHO international standards NIBSC code 05/132, when available |
|
Diagnostic specificity |
Unselected blood donors (including first-time donors)(1) |
≥5 000 |
≥99,5% |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total including the following specimens: positive for Borrelia burgdorferi sensu lato confirmed by IgG immunoblot; anti-HIV positive; RF+; other related microbial/infectious agents; systemic lupus erythematosus (SLE) patients; antiphospholipid antibody positive; pregnant women etc. |
Performance characteristic |
Specimens |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥300 positive specimens at different stages of the infection (early primary syphilis, secondary stage, and during late syphilis ) including high positive specimens, 50 low positive specimens, by at least two different serological tests (one of which is an enzyme immunoassay) for different antibodies to T.pallidum |
99% identification as “confirmed positive” or “indeterminate” |
Seroconversion panels |
At least 1 seroconversion panel, ≥1 if possible, including individual specimens from the early infection phase |
Diagnostic sensitivity during seroconversion shall represent the state of the art |
|
Analytical sensitivity |
Standards |
WHO international standards NIBSC code 05/132 |
|
Diagnostic specificity |
Blood donors |
≥200 |
≥ 99%; |
Clinical specimens |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 in total, including specimens from pregnant women and specimens with indeterminate results in other confirmatory assays. |
ANNEX XII
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF
TRYPANOSOMA CRUZI
INFECTION
Scope
Performance characteristic |
Specimens |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 positive specimens, including highly positive confirmed by at least two different serological tests for different antibodies to T.cruzi. Of those 400, ≥25 parasite positive specimens which have been confirmed by direct detection. |
99.5% overall sensitivity |
Seroconversion panels |
To be defined when available |
Diagnostic sensitivity during seroconversion shall represent the state of the art |
|
Analytical sensitivity |
Standards |
WHO international standards NIBSC code: 09/186 NIBSC code: 09/188 |
|
Diagnostic specificity |
Unselected donors (including first-time donors)(1) |
≥5 000 |
≥99,5% |
Hospitalised patients |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total including the following specimens: positive for anti-Toxoplasma gondii; at least 5 specimens positive for anti-Leishmania; RF+; related microbial agents or other infectious agents; SLE patients; antiphospholipid antibody positive patients; pregnant women, etc. |
Performance characteristic |
Specimens |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥300 positive specimens, including highly positive confirmed by at least two different serological tests for different antibodies to T.cruzi. Of those 300, ≥25 parasite positive specimens, which have been confirmed by direct detection. |
≥99% identification as “confirmed positive” or “indeterminate” |
Seroconversion panels |
As available |
Diagnostic sensitivity during seroconversion shall represent the state of the art, if applicable |
|
Analytical sensitivity |
Standards |
WHO international standards NIBSC code: 09/186 NIBSC code: 09/188 |
|
Diagnostic specificity |
Negative specimens |
≥200 |
≥99% |
Clinical specimens |
≥200 |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 in total, including specimens from pregnant women and specimens with indeterminate results in other confirmatory assays |
Table 3: NAT devices for T. cruzi DNA
Performance characteristic |
Specimens |
Specimen number, features, use |
Acceptance criteria |
Analytical sensitivity |
Characterized in-house reference preparation (as long as international reference materials are not available) |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device. LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).(2) |
According to the state of the art |
Diagnostic sensitivity: different T.cruzi strains / isolates |
Patient specimens from different regions determined as T.cruzi DNA positive by comparator device; sequence variants |
≥100 Dilution series of T.cruzi positive cell cultures (isolates) or T.cruzi positive materials from animal models may serve as potential substitutes |
According to the state of the art |
Diagnostic specificity |
Negative specimens |
≥100 |
According to the state of the art |
Cross-reactivity |
Potentially cross-reacting specimens |
≥10 human specimens positive for other parasites, e.g. Plasmodium species, Trypanosoma brucei. Positive cell cultures may serve as potential substitutes |
According to the state of the art |
Carry-over |
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The T.cruzi titres of the high positive specimens shall be representative of high T.cruzi titres occurring naturally. |
According to the state of the art |
Whole system failure rate |
|
≥100 T.cruzi DNA low-positive specimens shall be tested. These specimens shall contain a T.cruzi concentration equivalent to three times the 95 % positive cut-off T.cruzi concentration. |
≥99% positive |
ANNEX XIII
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF SEVERE ACUTE RESPIRATORY SYNDROME CORONAVIRUS 2 INFECTION
Scope
Performance characteristic |
Specimen |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥400 including specimens from early infection and post seroconversion(2) (within the first 21 days and after 21 days following the onset of symptoms); including specimens from asymptomatic or subclinical and mildly symptomatic (outpatient treatment) individuals; including specimens with low and high titers; including specimens from vaccinated individuals where appropriate(3); consideration of genetic variants |
≥90% sensitivity(4) for specimens taken >21 days after onset of symptoms(5); overall sensitivity including the early infection phase shall be at least equivalent to the comparator device(6) |
Seroconversion panels |
As far as available |
Seroconversion sensitivity comparable to other CE-marked devices |
|
Analytical sensitivity |
Reference preparations |
WHO International Standard (IS) for anti- SARS-CoV-2 (NIBSC code 20/136); WHO International Reference Panel (RP) for anti-SARS-CoV-2 antibodies (NIBSC codes 20/140, 20/142, 20/144, 20/148, 20/150) |
IS: for titre determinations / quantitative(7) result output; RP: all antibody assays |
Diagnostic specificity |
Negative specimens(8) |
≥400 specimens from non-infected and non-vaccinated individuals(9) |
>99% specificity(10) |
|
≥200 hospitalised patients (without SARS-CoV-2 infection) |
Potential limitations for specificity, if any, shall beidentified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total including RF+, pregnant women, specimens with antibodies against endemic human coronaviruses 229E, OC43, NL63, HKU1 and other pathogens of respiratory diseases such as influenza A, B, RSV etc. |
Performance characteristic |
Specimen |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥200(11) Specimens(12) with a significant proportion from the early phase of the infection (within 21 days after onset of symptoms) compared to specimens past seroconversion (>21 days after onset of symptoms); including specimens from asymptomatic, subclinical, mildly symptomatic (outpatient treatment) individuals; including freshly(13) vaccinated individuals if appropriate; consideration of genetic variants |
≥80% sensitivity(14) for specimens taken during the first 21 days after symptom onset(15); overall sensitivity shall be at least equivalent to the comparator device(16) of the same type (i.e. IgM and/or IgA) |
Seroconversion panels |
As far as available |
Seroconversion sensitivity comparable to other CE-marked devices |
|
Analytical sensitivity |
Standards |
N/A |
N/A |
Diagnostic specificity |
Negative specimens(17) |
≥200 specimens from non-infected and non-vaccinated individuals(18) |
≥98% specificity(19) |
|
≥100 from hospitalised patients (without SARS-CoV-2 infection) |
Potential limitations for specificity, if any, shall be identified |
|
Cross-reactivity |
Potentially cross-reacting specimens |
≥100 in total including RF+, pregnant women, specimens with antibodies against endemic human coronaviruses 229E, OC43, NL63, HKU1 and other pathogens of respiratory diseases such as influenza A, B, RSV etc. |
Performance characteristic |
Specimen |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥200 including specimens pre and post seroconversion (within the first 21 days and after 21 days following the onset of symptoms) |
Correct determination as “positive” (or “indeterminate”) |
Seroconversion panels/low titre panels |
as far as available |
||
Analytical sensitivity |
Standards |
N/A |
N/A |
Diagnostic specificity |
Negative specimens(21) |
≥200 from non-infected / non-vaccinated population |
No false positive results; correct determination as “negative” (or “indeterminate”) |
|
≥200 from hospitalised patients (without SARS-CoV-2 infection) |
||
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 in total including specimens with antibodies against endemic human coronaviruses 229E, OC43, NL63, HKU1 and other pathogens of respiratory diseases such as influenza A, B, RSV etc.; including specimens with indeterminate or false positive results in other anti-SARS-CoV-2 assays |
Performance characteristic |
Specimen |
Specimen number, features, use |
Acceptance criteria |
Diagnostic sensitivity |
Positive specimens |
≥100(22) NAT positive specimens(23) from early infection within the first 7 days after symptom onset(24); specimens shall represent naturally occurring viral loads(25); consideration of genetic variants(26); consideration of variations in specimen collection and/or specimen handling(27) |
Detection of >80% (rapid tests); detection of >85% (lab-based assays(28)); relative to SARS-CoV-2-NAT(29),(30) |
Analytical sensitivity |
Standards |
As soon as available |
Establishment of a LOD(31) |
Diagnostic specificity |
Negative specimens |
≥300 from non-infected individuals |
Specificity >98% (rapid tests) Specificity >99% (lab-based assays(28)) |
≥100 from hospitalised patients |
Potential limitations for specificity, if any, shall be identified |
||
Cross-reactivity |
Potentially cross-reacting specimens |
≥50 in total including virus-positive specimens of endemic human coronaviruses 229E, OC43, NL63, HKU1; influenza A, B, RSV, and other pathogens of respiratory diseases, eligible for differential diagnosis; including bacteria(32) present in the specimen taking area |
Performance characteristic |
Specimen |
SARS-CoV-2 RNA qualitative |
SARS-CoV-2 RNA quantitative |
Sensitivity |
|||
Analytical sensitivity: LOD |
WHO First International Standard SARS-CoV-2 RNA (NIBSC code 20/146; 7.70 Log10 IU/mL) Secondary standards calibrated against WHO IS |
According to Ph. Eur. NAT validation guideline: several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value |
According to Ph. Eur. NAT validation guideline: several dilution series of calibrated reference preparations into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value as LOD |
Quantification limit; quantification features |
WHO First International Standard SARS-CoV-2 RNA (NIBSC code 20/146; 7.70 Log10 IU/mL) Secondary standards calibrated against WHO IS |
|
Dilutions (half-log10 or less) of calibrated reference preparations; determination of lower, upper quantification limit, LOD, precision, accuracy, “linear” measuring range, “dynamic range”. Synthetic target nucleic acid may be used as secondary standard to achieve higher concentration levels. Reproducibility at different concentration levels to be shown |
Diagnostic sensitivity: different SARS-CoV-2 RNA strains |
Patient specimens determined as SARS-CoV-2 RNA positive by comparator device from different regions and outbreak clusters; sequence variants Dilution series of SARS-CoV-2 positive cell cultures (isolates) may serve as potential substitutes |
≥100(33) |
|
Quantification efficiency |
SARS-CoV-2 RNA positive patient specimens from different regions and outbreak clusters; sequence variants with quantitative values obtained by comparator device Dilution series of SARS-CoV-2 RNA positive cell cultures may serve as potential substitutes |
|
≥100 |
Inclusivity |
In silico analysis(34); at least two independent target gene regions in one test run (dual-target design) |
Evidence of suitable device design: primer/probe sequence alignments with published SARS-CoV-2 sequences |
Evidence of suitable device design: primer/probe sequence alignments with published SARS-CoV-2 sequences |
Specificity |
|||
Diagnostic specificity |
SARS-CoV-2 RNA negative human specimens |
≥500 |
≥100 |
In silico analysis(34) |
|
Evidence of suitable device design (sequence alignments); regular check of primer/probe sequences against sequence data bank entries |
Evidence of suitable device design evidence (sequence alignments); regular check of primer/probe sequences against sequence data bank entries |
Cross-reactivity |
specimens positive (various concentrations) for related human coronaviruses 229E, HKU1, OC43, NL63, MERS coronavirus; SARS CoV-1 if available; Influenza virus A, B; RSV; Legionella pneumophila; positive cell cultures may serve as potential substitutes |
≥20 in total |
≥20 in total |
Robustness |
|||
Carry-over |
|
At least 5 runs using alternating high positive and negative specimens. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. |
At least 5 runs using alternating high positive (known to occur naturally) and negative specimens |
Inhibition |
|
Internal control preferably to go through the whole NAT procedure |
Internal control preferably to go through the whole NAT procedure |
Whole system failure rate leading to false negative results: 99/100 assays positive |
|
≥100 specimens virus-spiked with 3 × the 95 % positive cut-off concentration (3 x LOD) |
≥100 specimens virus-spiked with 3 × the 95 % positive cut-off concentration (3 x LOD) |
Performance characteristic |
Specimens(36) |
Number of lay persons |
||||||||
Result interpretation(37) |
Interpretation of results(38) by lay persons reflecting the following range of reactivity levels:
|
≥100 |
||||||||
Diagnostic sensitivity(40) |
Lay persons that are known antigen positive(41) , (42) |
≥30 |
||||||||
Diagnostic specificity(43) |
Lay persons that do not know their status(39) |
≥60 |
Performance characteristic |
Specimens(45) |
Number of lay persons |
||||||||
Result interpretation(46) |
Interpretation of results(47) by lay persons reflecting the following range of reactivity levels:
|
≥100 |
||||||||
Diagnostic sensitivity(49) |
Lay persons that are known antibody positive(50) |
≥100 |
||||||||
Diagnostic specificity(51) |
Lay persons that do not know their status(48) |
≥100 |